We are a professional laboratory equipment manufacturer, glass reactor is one of our main products and include single layer glass reactor and double layer glass reactor.The capacity includes:2L glass reactor, 5L glass reactor, 10L glass reactor, 20L glass reactor, 50L glass reactor, 100L glass reactor, etc.
2.First, insert the pole into the corresponding aluminum seat on the heating bath to fix it.
3.Then fix the motor on the main pole, adjust the position and connect the stirring rod with a universal joint. The governor is inserted above the main pole, the five-core plug of the motor is connected to the five-core socket of the governor, and then the power is turned on; after starting the motor, slowly adjust the governor knob to see if the stirring rod of the motor is running smoothly. Adjust and calibrate the concentricity of the motor shaft and the stirring rod, adjust the stirring rod, and start operation after no swing.
4.Insert the glass fitting into the corresponding grinding port and fix it with the corresponding nut. Use the constant pressure dropping funnel clip to fix the constant pressure dropping funnel, and use the condenser clip to fix the condenser.
5.After adding a certain amount of water or oil to the bath, insert the cable plug on the right side of the bath into the three-wire socket.
6.Turn on the heating switch on the bath to set the temperature according to the heating requirements. Press the SET key to set the temperature, the upper row displays SP. Press the shift key << to make the number to be modified flash, press the up or down key to adjust the required value, and then press the SET key to return to the standard mode. Please use a thermometer to measure the precise water temperature directly, and the temperature control range is 0-400℃.
7. The power of the 10L glass reactor should be turned off in time after the reaction test is completed.